Nova 200 NanoLab: SEM & EDS: Difference between revisions

From the KNI Lab at Caltech
Jump to navigation Jump to search
 
(5 intermediate revisions by 3 users not shown)
Line 7: Line 7:
|RoomLocation = B203 Steele
|RoomLocation = B203 Steele
|LabPhone = 626-395-1542
|LabPhone = 626-395-1542
|PrimaryStaff = [[Annalena Wolff]]
|PrimaryStaff = [[Alireza Ghaffari]]
|StaffEmail = awolff@caltech.edu
|StaffEmail = alireza@caltech.edu
|StaffPhone = 626-395-5994
|StaffPhone = 626-395-3984
|Manufacturer = FEI (now Thermo Fisher)
|Manufacturer = FEI (now Thermo Fisher)
|Model = Nova 200 NanoLab
|Model = Nova 200 NanoLab
|Techniques = SEM, EDS<br>Immersion Lens Imaging
|Techniques = SEM, EDS<br>Immersion Lens Imaging
|RequestTraining = awolff@caltech.edu
|RequestTraining = alireza@caltech.edu
|EmailList = kni-sem-fib
|EmailList = kni-sem-fib
|EmailListName = SEM-FIB
|EmailListName = SEM-FIB
Line 32: Line 32:


== Resources ==
== Resources ==
===== Equipment Data =====
* [https://caltech.box.com/s/oy9kdmiwszygacoov2ko1a9njb8jghod Microscopy Pass-down equipment information]
===== SOPs & Manuals & SDS =====
===== SOPs & Manuals & SDS =====
* [https://caltech.box.com/s/mpsxkxmf5y8wjw9daijwbudeoqkkeudu KNI Microscopy Policies]
* [https://caltech.box.com/s/rpbtox8l31qi3kw3b014e3e8i4ctjpdy KNI Microscopy Policies]
* [https://caltech.box.com/s/4f1hqp83pwc1v1k5qke6xi68i28fjvz8 Nova200 and EDS SOP, Manuals, SDS]
* [https://caltech.box.com/s/4f1hqp83pwc1v1k5qke6xi68i28fjvz8 Nova200 and EDS SOP, Manuals, SDS]
* [https://caltech.box.com/s/og4309108q4k2jwhkaxqtpiujg2al5iu Nova NanoLab Operation Manual]
* [https://caltech.box.com/s/og4309108q4k2jwhkaxqtpiujg2al5iu Nova NanoLab Operation Manual]
Line 56: Line 54:
* [https://caltech.box.com/s/ijd8gprg9gcavb6of5uegu7osinzsdet Guide to Optimizing SEM Imaging]
* [https://caltech.box.com/s/ijd8gprg9gcavb6of5uegu7osinzsdet Guide to Optimizing SEM Imaging]
===== Presentations =====
===== Presentations =====
* SEM Essentials: Capabilities and Limitations of Scanning Electron Microscopy; and SEM Setup, Parameters and Theory for Successful Operations and Measurements
**[https://caltech.box.com/s/eg20eqw7si5ka0bms251qurfwhys0hta Download .pptx Slides]
* Scanning Electron Microscopy: Principles, Techniques & Applications
* Scanning Electron Microscopy: Principles, Techniques & Applications
** [https://caltech.box.com/s/lulkj0pwm053akyya1shazg8wzgudq9f Pptx Slides] | [https://youtu.be/Zh21tp3aPEw YouTube Lecture]
** [https://caltech.box.com/s/lulkj0pwm053akyya1shazg8wzgudq9f .pptx Slides] | [https://youtu.be/Zh21tp3aPEw YouTube Lecture]
* Gallium Focused Ion Beam Microscopy: Principles, Techniques & Applications
* Gallium Focused Ion Beam Microscopy: Principles, Techniques & Applications
** [https://caltech.box.com/s/f4k8jan85n5lf6f2tutjx4rkfzjq7y68 PPtx Slides] | [https://youtu.be/3eSzisbNcGo YouTube Lecture]
** [https://caltech.box.com/s/f4k8jan85n5lf6f2tutjx4rkfzjq7y68 .pptx Slides] | [https://youtu.be/3eSzisbNcGo YouTube Lecture]
 


===== Simulation Software =====
===== Simulation Software =====

Latest revision as of 22:35, 5 February 2025

Nova 200
Nova-200-NanoLab.jpg
Instrument Type Microscopy
Techniques SEM, EDS
Immersion Lens Imaging
Staff Manager Alireza Ghaffari
Staff Email alireza@caltech.edu
Staff Phone 626-395-3984
Reserve time on FBS
Request training via FBS User Dashboard
Lab Location B203 Steele
Lab Phone 626-395-1542
Manufacturer FEI (now Thermo Fisher)
Model Nova 200 NanoLab

Description

The Nova 200 is the KNI's highest-resolution field-emission gun (FEG) analytical scanning electron microscope (SEM), equipped with an immersion lens for imaging sub-10 nm features and energy dispersive spectroscopy (EDS) for compositional analysis. It is also outfitted with a gallium focused ion beam (Ga-FIB) column, which is currently not operational because the Nova 600 NanoLab and ORION NanoFab together meet the KNI's Ga-FIB demand; Ga-FIB could be reactivated on the Nova 200 in the future. See a full list of training and educational resources for this instrument below.

SEM Applications
  • Ultra-High-Resolution Imaging (Immersion Mode aka UHR Mode)
  • High-Resolution Imaging (Field-Free Mode aka Normal Mode)
  • Secondary Electron (SE) imaging with an Everhart-Thornley Detector (ETD) & Through-the-Lens Detector (TLD)
  • Backscattered Electron (BSE) imaging with a TLD
  • Tungsten deposition via Gas Injection System (GIS)
  • Automated imaging with RunScript program & AutoScript language
EDS Applications
  • Spectrum acquisition for qualitative and quantitative compositional analysis
  • Linescan acquisition for 1D spatial compositional analysis
  • Map acquisition for 2D spatial compositional analysis

Resources

SOPs & Manuals & SDS
Video Tutorials
Graphical Handouts
Presentations
  • SEM Essentials: Capabilities and Limitations of Scanning Electron Microscopy; and SEM Setup, Parameters and Theory for Successful Operations and Measurements
  • Scanning Electron Microscopy: Principles, Techniques & Applications
  • Gallium Focused Ion Beam Microscopy: Principles, Techniques & Applications
Simulation Software
Calibrate Measurements with NIST Standard
  • The KNI has a NIST-traceable standard against which SEM measurements can be compared. See Slides 54-55 of the SEM Presentation for details. Ask staff for help finding and using the standard in the lab.
Sample Preparation
  • Use the Carbon Evaporator to make non-conductive samples conductive by applying 2-10 nm of evaporated carbon.
  • Use the O2/Ar Plasma Cleaner to remove hydrocarbons from the sample surface to avoid creating dark contamination spots on your features while imaging them.
Stubs for specimen mounting
  • Stubs used for mounting specimens are considered a personal, consumable item in the KNI. There are some stubs at each Microscope which can be used by any KNI microscopy user. You can also buy your own stubs so that you can keep them clean and available to you. There are many stub geometries and configurations. If you chose to buy your own stubs, please show them to the staff microscopist prior to using them: some stubs including stubs with copper clips have large height differences and can only be used safely in specific operating conditions.
Guide to Choosing KNI SEMs & FIBs

Specifications

Manufacturer Specifications
SEM Specifications
  • Minimum Feature Size Resolved in Immersion Mode: ~5 nm
  • Voltage Range: 0.2 to 30.0 kV
  • Current Range: ~10 pA to 20 nA
  • Apertures: 30 μm, 40 μm, 50 μm, 100 μm
  • Eucentric Height: ~4.8 mm working distance (WD)
  • Stage Range: ±25 mm X & Y travel, 50 mm Z travel, -12 to 58° tilt, 360° rotation
  • ETD Grid Bias Range: -150 to 300 V
  • TLD Bias Range: -100 to 150 V
  • Ultimate Vacuum: 5e-6 mbar