TEM Sample Preparation Equipment: Difference between revisions
Jump to navigation
Jump to search
mNo edit summary |
|||
(4 intermediate revisions by 3 users not shown) | |||
Line 7: | Line 7: | ||
|RoomLocation = B242A Keck | |RoomLocation = B242A Keck | ||
|LabPhone = 626-395-8908 | |LabPhone = 626-395-8908 | ||
|PrimaryStaff = [[ | |PrimaryStaff = [[Guy A. DeRose, PhD]] | ||
|StaffEmail = | |StaffEmail = derose@caltech.edu | ||
|StaffPhone = 626-395- | |StaffPhone = 626-395-3423 | ||
|Manufacturer = Assorted (see list on page) | |Manufacturer = Assorted (see list on page) | ||
|Model = Assorted (see list on page) | |Model = Assorted (see list on page) | ||
|Techniques = Cross-Section Sample Prep,<br>Sample Polishing,<br>3 mm Core Drilling,<br>Dimpling, Argon Milling | |Techniques = Cross-Section Sample Prep,<br>Sample Polishing,<br>3 mm Core Drilling,<br>Dimpling, Argon Milling | ||
|RequestTraining = | |RequestTraining = derose@caltech.edu | ||
|EmailList = kni-tf30 | |EmailList = kni-tf30 | ||
|EmailListName = TF-30 | |EmailListName = TF-30 | ||
Line 29: | Line 29: | ||
[[Image:Dimpler-and-Disk-Cuttter.jpg|thumb|top|upright=1.0|The Dimpler (left) and 3 mm Disk Cutter (right)]] | [[Image:Dimpler-and-Disk-Cuttter.jpg|thumb|top|upright=1.0|The Dimpler (left) and 3 mm Disk Cutter (right)]] | ||
[[Image:Polishing-Station-Allied-TechPrep.jpg|thumb|top|upright=1.0|The polishing station with automated handling of the polishing jig and control of its sweeping motion]] | [[Image:Polishing-Station-Allied-TechPrep.jpg|thumb|top|upright=1.0|The polishing station with automated handling of the polishing jig and control of its sweeping motion]] | ||
* This area is currently under refurbishment and not available. Please contact Annalena Wolff for more information | |||
* Low-Angle Argon Milling & Polishing Station: Fischione Model 1010 (shown in infobox image) | * Low-Angle Argon Milling & Polishing Station: Fischione Model 1010 (shown in infobox image) | ||
* Dimpler: Fischione Model 2000 | * Dimpler: Fischione Model 2000 | ||
Line 47: | Line 48: | ||
===== Videos ===== | ===== Videos ===== | ||
* [https://www.youtube.com/watch?v=NyxfiC13kQ8 Ultrasonic Disk Cutter Model 170 Operation] | * [https://www.youtube.com/watch?v=NyxfiC13kQ8 Ultrasonic Disk Cutter Model 170 Operation] | ||
<br> | |||
<br> | |||
== Related Instrumentation in the KNI == | |||
===== Transmission Electron Microscopes ===== | |||
* [[Tecnai TF-30: 300 kV TEM, STEM, EDS & HAADF | Tecnai TF-30: TEM, STEM, EDS & HAADF (50-300 kV)]] | |||
<!--- | |||
* [[Tecnai TF-20: 200 kV TEM, STEM, EDS, EELS, EFTEM & Lithography | Tecnai TF-20: TEM, STEM, EDS, EELS, EFTEM & Lithography (40-200 kV)]] | |||
---> | |||
===== Sample Preparation for TEM ===== | |||
* [[Nova 600 NanoLab: SEM, Ga-FIB, GIS & Omniprobe|Nova 600 NanoLab: SEM, Ga-FIB, GIS & Omniprobe]] | |||
* [[TEM Sample Preparation Equipment | TEM Sample Preparation Equipment: Polishing Stations, 3 mm Disk Cutter, Dimpler, Argon Ion Mill]] | |||
===== Sample Preparation for Microscopy ===== | |||
* [[Carbon Evaporator | Carbon Evaporator (Leica EM ACE600) to make samples conductive]] | |||
* [[Tergeo Plus ICP- & CCP-RIE: Oxygen & Argon Plasma Cleaner | Oxygen & Argon Plasma Cleaner (Tergeo Plus ICP- & CCP-RIE) to remove hydrocarbons from surface]] |
Latest revision as of 18:04, 29 June 2023
|
Description
The KNI offers a lab in the Keck sub-basement for preparing TEM samples by traditional means, often to produce cross-section specimens from a bulk sample. This typically involves the following steps:
- Glue together a stack of materials, e.g. with the surface of two samples facing each other
(if the surface is what you want to observe in the TEM) - Cut the stack into thin slices of material, in the appropriate orientation
(the thin dimension here will be the same thin dimension in the TEM) - Cut out a 3 mm circle from this thin slice using a Disk Cutter
- Polish that thin disk down to a thickness of ~100 μm
- Use a Dimpler to polish a hemispherical dimple into the middle of the disk,
such that only a few microns of material remain at the thinnest portion - Use an Argon Mill to perform the final polish of the thinnest portion,
until a hole has been bored through the middle and
the edges of that hole are thin enough to be electron-transparent in the TEM
List of Equipment Available
- This area is currently under refurbishment and not available. Please contact Annalena Wolff for more information
- Low-Angle Argon Milling & Polishing Station: Fischione Model 1010 (shown in infobox image)
- Dimpler: Fischione Model 2000
- Ultrasonic Disk Cutter: Model 170
- Polishing Station: Allied TechPrep
- Polishing Station: Buehler Ecomet 3
- Low Speed Saw: Isomet
- Roll Grinder: Buehler Handimet 2
- Various polishing media & accessories:
- Media: grit paper, lapping films, polishing cloths, diamond suspensions, alumina poweders
- Accessories: tripod polisher (South Bay Technology), endpoint polisher (Fischione Model 160)
Resources
SOPs
- Paper copies of SOPs are available in the lab for the Argon Ion Mill & Dimpler
Manufacturer Manuals
Videos
Related Instrumentation in the KNI
Transmission Electron Microscopes
Sample Preparation for TEM
- Nova 600 NanoLab: SEM, Ga-FIB, GIS & Omniprobe
- TEM Sample Preparation Equipment: Polishing Stations, 3 mm Disk Cutter, Dimpler, Argon Ion Mill